Abstract:
Background: Automated full blood cell count (FBC) is a common test. The precision of FBC is 
monitored by internal quality control (IQC). Manual visual counting of cells in blood smears using 
light microscopes (LM) is an IQC method. Conversion factors calculated by visual counting cells 
in blood smears are vary between cell types and LM due to the differences in field of view (FOV). 
Due to differences in FOV, one conversion factor cannot be used between different LM brands 
and types. 
Objective: To determine conversion factors for automated cell counts estimated from cell counts 
in blood smears using different LM 
Methods: EDTA blood samples (n=30) and their FBC results by Mindray BC 5150 analyzer were 
obtained. Manual cell counting was performed using Leishman-stained blood smears on Olympus 
CX 23, CX 31, CX 43, Euromax, LaboMed, and Carl Zeiss microscopes. Red cells and platelets in 
ten oil immersion fields and white cells in ten high-power fields were counted to obtain mean 
manual cell counts for each sample. Duplicate counting was done to eliminate any personal errors. 
The conversion factors were calculated by dividing automated cell count by the corresponding 
manual mean count. 
Results: The conversion factor for platelet count was 13 000 for all except the Carl Zeiss 
microscope (12 000). The conversion factor for red blood cells was 1.4 for all except the LaboMed 
microscope (1.3). The conversion factor for total white blood cell count was 4000 for all the six 
microscopes.  
Conclusions: This study generated conversion factors for different microscopes and cell types for 
Mindray BC 5150 analyzer. Further studies are needed to evaluate the application of these 
conversion factors.